Title 21: Food and Drugs
« Home / Title 21 - Food and Drugs--Volume 8 / CHAPTER I - FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) / SUBCHAPTER H - MEDICAL DEVICES
/ PART 876 - GASTROENTEROLOGY-UROLOGY DEVICES
/ Subpart F - Therapeutic Devices / § 876.5955 Peritoneo-venous shunt.
§ 876.5955 Peritoneo-venous shunt.
(a) Identification. A peritoneo-venous shunt is an implanted device that consists of a catheter and a pressure activated one-way valve. The catheter is implanted with one end in the peritoneal cavity and the other in a large vein. This device enables ascitic fluid in the peritoneal cavity to flow into the venous system for the treatment of intractable ascites.
(b) Classification. Class II. The special controls for this device are FDA's:
(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices - Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) Backflow specification and testing to prevent reflux of blood into the shunt.